DRUCK Pressure Indicator
DRUCK Pressure Indicator
Model: DPI705
Brand: druck
Origin: UK
概要
- ±1.69 psi to 20,000 psi pressure range
- 0.05% full scale accuracy
- Integral calibration record
- Leak test, tare, maximum/minimum and filter
- Hazardous Area (IS) version available
The Druck DPI 705E Pressure Indicator combine tough and rugged design with accurate and reliable measurements. The DPI 705E Series is designed for single handed operation and provides many essential features required for routine maintenance and system troubleshooting.
The 705E features 48 pressure ranges from ±1.69 psi to 20,000 psi (±25 mbar to 1,400 bar / 2.5 KPa to 140 MPa). The leak test is used to determine if there is a leak in the system by recording the pressure change over of time. The leak test can also be used with the RTD sensor to record a temperature change over time. It is user defined as 1, 3 or minute test.
- Two Models:
- DPI705E ===> Safe Area pressure indicator
- DPI705EIS ==> Hazardous Area pressure indicator
- PRESSURE RANGES (both models):
- 0 to 10" WC through to 20,000 psi (0 to 25 mbar through to 1400 bar)
- PRESSURE ACCURACY (calibration record included):
- Standard Accuracy => 0.1% full scale (FS)
- High Accuracy ====> 0.05% full scale (FS)
- Optional remote plug and play pressure and Resistance Temperature Detector (RTD) sensors
- Optional pneumatic and hydraulic hand pumps
Accuracy Options
SKU | Accuracy Level | Accuracy |
1 | Standard | ±0.1% |
2 | High | ±0.05% |
3 | Premium | ±0.025% |
Internal Pressure Sensor Options
Pressure Range | Abs | Gauge | Differential | Barometric | Standard Accuracy | High Accuracy | Precise Accuracy |
25 mbar / 10 inH2O / 2.5 kPa | - | 008G | 008L | - | G, L | - | - |
70 mbar / 1 psi / 7 kPa | - | 01G | 01L | - | G, L | - | - |
200 mbar / 3 psi / 20 kPa | - | 02G | 02L | - | G, L | - | - |
350 mbar / 5 psi / 35 kPa | 03A | 03G | 03L | - | G, A, L | G, L | - |
700 mbar / 10 psi / 70 kPa | 04A | 04G | 04L | - | G, A, L | G, L | - |
1 bar / 15 psi / 100 kPa | 05A | 05G | 05L | - | G, A, L | G, A, L | - |
75-115 kPa (Barometric) | - | - | - | 05B | B | B | - |
2 bar / 30 psi / 200 kPa | 07A | 07G | 07L | - | G, A, L | G, A, L | G,L |
3.5 bar / 50 psi / 350 kPa | 08A | 08G | - | - | G, A | G, A | G |
7 bar / 100 psi / 700 kPa | 10A | 10G | - | - | G, A | G, A | G |
10 bar / 150 psi / 1000 kPa | 11A | 11G | - | - | G, A | G, A | G, A |
20 bar / 300 psi / 2 MPa | 13A | 13G | - | - | G, A | G, A | G, A |
35 bar / 500 psi / 3.5 MPa | 14A | 14G | - | - | G, A | G, A | G, A |
70 bar / 1000 psi / 7 MPa | 16A | 16G | - | - | G, A | G, A | G, A |
100 bar / 1500 psi / 10 MPa | 165A | 165G | - | - | G, A | G, A | G, A |
135 bar / 2000 psi / 13.5 MPa | 17A | 17G | - | - | G, A | G, A | G, A |
200 bar / 3000 psi / 20 MPa | 18A | 18G | - | - | G, A | G, A | G, A |
350 bar / 5000 psi / 35 MPa | 20A | - | - | - | A | A | A |
700 bar / 10000 psi / 70MPa | 22A | - | - | - | A | A | A |
Note | General | Do not apply pressure greater than 1 bar (16 psi) absolute to the negative port of differential pressure indicators. | |||||
Pressure Connections | 25 mbar – 200 bar (10 inH2O - 3000 psi / 2.5 kPa - 20 MPa): G1/8 BSP female | ||||||
Differential reference port: Legris 4 mm tubing adaptor | |||||||
≥350 bar (5000 psi / 35 MPa): Autoclave 9/16 x 18 UNF Male | |||||||
Key | G = Gauge | ||||||
L = Differential | |||||||
A = Absolute | |||||||
B = Baro |
Pressure Fitting Options
SKU | Pressure Fitting | Notes |
P1 | G1/8 BSP Female | For ranges less than 350 bar |
P2 | G1/4 BSP Female Adaptor | For ranges less than 350 bar |
P3 | 1/8 NPT Female Adaptor | For ranges less than 350 bar |
P4 | 1/4 NPT Female Adaptor | For ranges less than 350 bar |
P5 | Quick-Fit Adaptor | For ranges less than 350 bar |
P6 | 9/16 x 18 UNF Male | Mandatory for ranges greater than/equal to 350 bar |
Pressure Unit Options
SKU | Pressure Units |
U0 | All Pressure units (Default selection) |
U1 | Pa (Si) Pressure units only |
Additional Options
SKU | Option |
OP0 | No option required |
OP1 | Negative calibration for gauge ranges less than 35 bar (If selected, the calibration certificate includes values down to -1 bar g.) |
応用
The Druck DPI 705E Pressure Indicator combine tough and rugged design with accurate and reliable measurements. The DPI 705E Series is designed for single handed operation and provides many essential features required for routine maintenance and system troubleshooting.
The 705E features 48 pressure ranges from ±1.69 psi to 20,000 psi (±25 mbar to 1,400 bar / 2.5 KPa to 140 MPa). The leak test is used to determine if there is a leak in the system by recording the pressure change over of time. The leak test can also be used with the RTD sensor to record a temperature change over time. It is user defined as 1, 3 or minute test.
- Two Models:
- DPI705E ===> Safe Area pressure indicator
- DPI705EIS ==> Hazardous Area pressure indicator
- PRESSURE RANGES (both models):
- 0 to 10" WC through to 20,000 psi (0 to 25 mbar through to 1400 bar)
- PRESSURE ACCURACY (calibration record included):
- Standard Accuracy => 0.1% full scale (FS)
- High Accuracy ====> 0.05% full scale (FS)
- Optional remote plug and play pressure and Resistance Temperature Detector (RTD) sensors
- Optional pneumatic and hydraulic hand pumps
仕様
Accuracy Options
SKU | Accuracy Level | Accuracy |
1 | Standard | ±0.1% |
2 | High | ±0.05% |
3 | Premium | ±0.025% |
Internal Pressure Sensor Options
Pressure Range | Abs | Gauge | Differential | Barometric | Standard Accuracy | High Accuracy | Precise Accuracy |
25 mbar / 10 inH2O / 2.5 kPa | - | 008G | 008L | - | G, L | - | - |
70 mbar / 1 psi / 7 kPa | - | 01G | 01L | - | G, L | - | - |
200 mbar / 3 psi / 20 kPa | - | 02G | 02L | - | G, L | - | - |
350 mbar / 5 psi / 35 kPa | 03A | 03G | 03L | - | G, A, L | G, L | - |
700 mbar / 10 psi / 70 kPa | 04A | 04G | 04L | - | G, A, L | G, L | - |
1 bar / 15 psi / 100 kPa | 05A | 05G | 05L | - | G, A, L | G, A, L | - |
75-115 kPa (Barometric) | - | - | - | 05B | B | B | - |
2 bar / 30 psi / 200 kPa | 07A | 07G | 07L | - | G, A, L | G, A, L | G,L |
3.5 bar / 50 psi / 350 kPa | 08A | 08G | - | - | G, A | G, A | G |
7 bar / 100 psi / 700 kPa | 10A | 10G | - | - | G, A | G, A | G |
10 bar / 150 psi / 1000 kPa | 11A | 11G | - | - | G, A | G, A | G, A |
20 bar / 300 psi / 2 MPa | 13A | 13G | - | - | G, A | G, A | G, A |
35 bar / 500 psi / 3.5 MPa | 14A | 14G | - | - | G, A | G, A | G, A |
70 bar / 1000 psi / 7 MPa | 16A | 16G | - | - | G, A | G, A | G, A |
100 bar / 1500 psi / 10 MPa | 165A | 165G | - | - | G, A | G, A | G, A |
135 bar / 2000 psi / 13.5 MPa | 17A | 17G | - | - | G, A | G, A | G, A |
200 bar / 3000 psi / 20 MPa | 18A | 18G | - | - | G, A | G, A | G, A |
350 bar / 5000 psi / 35 MPa | 20A | - | - | - | A | A | A |
700 bar / 10000 psi / 70MPa | 22A | - | - | - | A | A | A |
Note | General | Do not apply pressure greater than 1 bar (16 psi) absolute to the negative port of differential pressure indicators. | |||||
Pressure Connections | 25 mbar – 200 bar (10 inH2O - 3000 psi / 2.5 kPa - 20 MPa): G1/8 BSP female | ||||||
Differential reference port: Legris 4 mm tubing adaptor | |||||||
≥350 bar (5000 psi / 35 MPa): Autoclave 9/16 x 18 UNF Male | |||||||
Key | G = Gauge | ||||||
L = Differential | |||||||
A = Absolute | |||||||
B = Baro |
Pressure Fitting Options
SKU | Pressure Fitting | Notes |
P1 | G1/8 BSP Female | For ranges less than 350 bar |
P2 | G1/4 BSP Female Adaptor | For ranges less than 350 bar |
P3 | 1/8 NPT Female Adaptor | For ranges less than 350 bar |
P4 | 1/4 NPT Female Adaptor | For ranges less than 350 bar |
P5 | Quick-Fit Adaptor | For ranges less than 350 bar |
P6 | 9/16 x 18 UNF Male | Mandatory for ranges greater than/equal to 350 bar |
Pressure Unit Options
SKU | Pressure Units |
U0 | All Pressure units (Default selection) |
U1 | Pa (Si) Pressure units only |
Additional Options
SKU | Option |
OP0 | No option required |
OP1 | Negative calibration for gauge ranges less than 35 bar (If selected, the calibration certificate includes values down to -1 bar g.) |