SUPER MEGOHM METER
SUPER MEGOHM METER
Model: SM7110
Brand: Hioki
Origin: Japan
Summary
Rapid 6.4ms Super Megohm and Low Current Testing
The Hioki SM Series Super Megohm Meters are high resistance and ultra low current inspection devices with a 0.1 fA resolution to offer precise testing on production lines. The SM7110 is designed for use in evaluating and testing the performance of electronic components used in products such as smartphones and hybrid vehicles, and supports measurement voltage output of up to 1000 V.
Number of channels | 1 ch | |||||
---|---|---|---|---|---|---|
DC current measurement | 20 pA range (0.1 fA resolution), Accuracy: ±(2.0 % of rdg. +30 dgt.) 200 pA range (1.0 fA resolution), Accuracy: ±(1.0 % of rdg. +30 dgt.) 2 nA range (10 fA resolution), Accuracy: ±(0.5 % of rdg. +20 dgt.) 20 nA range (100 fA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 nA range (1 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 2 μA range (10 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 20 μA range (100 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 μA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) *2 mA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +30 dgt.) (1) Measurement speed SLOW2 (internal integration time 13PLC) (2) At a temperature of 23 °C ±5 °C with humidity of 85% rh (3) 2 mA range (Measurement speed FAST only) |
|||||
Resistance measurement capabilities | 1 × 10 ^3 Ω to 2 × 10 ^19 Ω Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy. |
|||||
Setting voltage range (Accuracy) |
0.1 to 100.0 V, 100 mV resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. 100.1 to 1000 V, 1 V resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. |
|||||
Current Limiter | 0.1 to 250.0 V: 5/ 10/ 50 mA, 251 to 1000 V: 5/ 10 mA, to 2000 V:1.8 mA | |||||
Measurement time setting | Delay: 0 to 9,999 ms | |||||
Functions | Comparator, averaging, self-calibration, jig Capacity open correction, cable length correction, surface resistivity, volume resistivity, voltage monitor, contact check | |||||
Program function | 10 types of discharge, charge, measure and measurement sequence discharge patterns can be programmed. | |||||
Display | LCD (8 lines of 30 characters), with backlight, High voltage warning indicator | |||||
Interfaces | USB, RS-232C, GP-IB, EXT I/O (NPN/PNP can be switched) | |||||
Power supply | 100 to 240V AC , 50/60 Hz, 45 VA | |||||
Dimensions and mass | 330 mm (12.99 in)W × 80 mm (3.15 in)H × 450 mm (17.72 in)D, 5.9 kg (208.1 oz) | |||||
Included accessories | Power cord ×1, Instruction manual ×1, CD-R (Communications command instruction manual, USB driver) ×1, EXT I/O male connector ×1, Short plug ×1 |
Application
The Hioki SM Series Super Megohm Meters are high resistance and ultra low current inspection devices with a 0.1 fA resolution to offer precise testing on production lines. The SM7110 is designed for use in evaluating and testing the performance of electronic components used in products such as smartphones and hybrid vehicles, and supports measurement voltage output of up to 1000 V.
Specifications
Number of channels | 1 ch | |||||
---|---|---|---|---|---|---|
DC current measurement | 20 pA range (0.1 fA resolution), Accuracy: ±(2.0 % of rdg. +30 dgt.) 200 pA range (1.0 fA resolution), Accuracy: ±(1.0 % of rdg. +30 dgt.) 2 nA range (10 fA resolution), Accuracy: ±(0.5 % of rdg. +20 dgt.) 20 nA range (100 fA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 nA range (1 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 2 μA range (10 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 20 μA range (100 pA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) 200 μA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +10 dgt.) *2 mA range (1 nA resolution), Accuracy: ±(0.5 % of rdg. +30 dgt.) (1) Measurement speed SLOW2 (internal integration time 13PLC) (2) At a temperature of 23 °C ±5 °C with humidity of 85% rh (3) 2 mA range (Measurement speed FAST only) |
|||||
Resistance measurement capabilities | 1 × 10 ^3 Ω to 2 × 10 ^19 Ω Note: Resistance measurement accuracy is defined by the current range accuracy and voltage setting accuracy. |
|||||
Setting voltage range (Accuracy) |
0.1 to 100.0 V, 100 mV resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. 100.1 to 1000 V, 1 V resolution, Accuracy: ±0.1 % of setting ±0.05% f.s. |
|||||
Current Limiter | 0.1 to 250.0 V: 5/ 10/ 50 mA, 251 to 1000 V: 5/ 10 mA, to 2000 V:1.8 mA | |||||
Measurement time setting | Delay: 0 to 9,999 ms | |||||
Functions | Comparator, averaging, self-calibration, jig Capacity open correction, cable length correction, surface resistivity, volume resistivity, voltage monitor, contact check | |||||
Program function | 10 types of discharge, charge, measure and measurement sequence discharge patterns can be programmed. | |||||
Display | LCD (8 lines of 30 characters), with backlight, High voltage warning indicator | |||||
Interfaces | USB, RS-232C, GP-IB, EXT I/O (NPN/PNP can be switched) | |||||
Power supply | 100 to 240V AC , 50/60 Hz, 45 VA | |||||
Dimensions and mass | 330 mm (12.99 in)W × 80 mm (3.15 in)H × 450 mm (17.72 in)D, 5.9 kg (208.1 oz) | |||||
Included accessories | Power cord ×1, Instruction manual ×1, CD-R (Communications command instruction manual, USB driver) ×1, EXT I/O male connector ×1, Short plug ×1 |